The instrument standard IEC 61043 specifies the requirements for phase matching in terms of the pressure-intensity residual index δpI0 [dB], which is the FpI when the probe microphones are exposed to an identical pink noise sound field. The required δpI0 is defined based on a phase-match requirement for the whole analyser.
To obtain the δpI0, the Nor150 is delivered with the Nor1294 Phase Verification Coupler. This coupler ensures an identical sound field for both microphones, with exceptional low-frequency response.
Traditionally, the analyser and the probe is calibrated separately and as a total in a laboratory to ensure the requirement is fulfilled. This is still a requirement, but using the Nor150 and the Nor1294 together, both verification and a new correction algorithm is available to the user. A much better Dynamic Capability may then be obtained.
A phase correction algorithm has two main benefits:
1. A higher Ld allows highly reactive fields to be measured. This means that higher levels of extraneous noise may be separated from the noise source of interest.
2. A single spacer may be used for the whole frequency range of interest. The 12 mm spacer with phase correction applied can measure from 31.5 Hz and up to 6.3 kHz.