In Advanced View, different functions are avalailable depending on the scope and standard.
Press to toggle between different Advanced View functions.
The weighted sum frequency range is indicated with white background, bands outside the range is gray.
The different views are described pr standard below:
Use the cursor keys or tap a frequency band to get the values for the selected band.
Tap and drag vertically on the graph to change the visible range.
Press and hold anywhere on the graph to bring up the context menu:
Change the Y-axis resolution.
Toggle pre-weighting of the spectrum using the A-weight. APRW is indicated in the top-right corner of the screen.
Excludes or includes the selected band from the calculation. Indicated by a blue mark/icon.
In Simple and Advanced View, the network bands A, C and Z are calculated from the Weighted Sum Freq. Range. Thus, bands outside this range is not included in the calculation, even when though are displayed in the spectrum.
Each band may have a warning indicator. Depending on the selected standard, different warnings apply.
The A/C/Z bands have summary warnings for all bands within the Weighted Sum Frequency Range, and also special warnings applying to the total.
Band warnings are indicated by coloured exclamation marks in the bottom of the graph view.
Cursor position warning is indicated by a triangle between the measurement functions and values above the graph.
A warning representing the total (A/C/Z) of the measurement is presented in the upper-right corner of the screen.
Press to show bring up the Warning Panel with detailed information about cursor position warnings.
Press to toggle between spectrum and tabular view.
Press to toggle between Simple View and Advanced View. If reviewing a Surface measurement or the Total, you will also toggle through the corresponding Scope view.
Tap to step back to the Scope view.
Setup NorRemote:
Enable NorRemote