This page will guide you through the setup of the Nor1290 in the Nor150. Other probes will follow the same strategy.
Enter Transducers menu

To enter the Transducers menu, follow this sequence:

SETUP



Add Microphones

Start by tapping

Add

then the upper


and select


Accept, then tap into the lower


Select the "40GI", which is the individual mic type.

Give the microphonse a name, such as "Nor1290 A". Then fill in the rest of the information: Serial numbers, Laboratory, date and verified sensitivity as found in the calibration chart.

Accept the configuration, then repeat the procedure for the second mic.

Add the Intensity probe

In the transducer meny, tap

Add

once more. This time, select "Intensity Probe" as the type. Give it a name, and select your newly created microphones as the input. Channel 1 should have mic A connected.

Enter the serial number and accept the new probe configuration.

Set the probe as input

To use the probe, press

SETUP



Tap


to find the new probe. Select it, then accept.

You are now ready to calibrate and start measuring.

Sound Intensity help index